BRONKHORST®
Page29DeviceNetinterface 9.17.026
5 EDS‐FILE
ToassistusersinconfiguringthePolledI/Ocommunication,anElectronicsDataSheet(EDS)filewillbeprovided.Each
typeDeviceNetinstrumentshouldhaveitsownEDS‐filewithinstrumentspecificationstotellthemasterconfiguration
softwarewhichfacilities/featurestheinstruments/slavesoffertothemaster.ForBronkhorstmeter/controllerthefile
iscalled:BHT_DMFC.EDS.ThisfileisavailableontheMultibusdocumentation/softwaretoolCD.
ThisEDS‐fileisatext‐filecontaining:
Identificationinfo:
Modelname: “Bronkhorstmeter/controller”
Vendorname: “BronkhorstHigh‐TechB.V.”
VendorID: 706
Settinginfo:
IOCharacteristicssection,containsavailablepolledI/O
combinations(IOAssemblyInstances)
Parametersection,theconfigurationsoftwareusesthissectiontogenerateadialogboxwheretheusercanenter
instrumentparametersforcustomising(seeparagraph2.5.2).
Afterstarting‐upyourmasterconfigurationsoftware,thisEDS‐fileshouldbeload/import/copied.Thisisneededonly
once(untila
eventualnextrevisionfromthefile).