Bronkhorst Multibus Bedienungsanleitung Seite 8

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 50
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 7
BRONKHORST®
Page 8 Operational instructions for digital multibus instruments 9.17.023
Digital instruments can be operated by means of:
1. Analog interface. (0...5Vdc/0...10Vdc/0...20mA/4...20mA)
2. RS232 interface (connected to COM-port by means of special cable (Default speed 38400 Baud)
3. FLOW-BUS
4. PROFIBUS-DP
5. DeviceNet
6. Modbus
7. EtherCAT
Option 1 and 2 are always present on Multibus instruments. Option 3, 4, 5 and 6 are optional. Operation via analog
interface, RS232 interface and an optional field bus can be performed at the same time. A special parameter called
“control mode” indicates to which setpoint the controller should listen: analog or digital (via field bus or RS232). The
RS232 interface behaves like a FLOW-BUS interface. When using more digital interfaces at the same time, reading can
be done simultaneously without problems. When changing a parameter value, the last value send by an interface will
be valid.
Also the micro push-button switch and the LED’s on top of the instrument can be used for manual operation of some
options.
The green LED will indicate in what mode the instrument is active.
The red LED will indicate info / error / warning situations.