Doc. no.: 9.17.084A Date: 17-06-2015Instruction ManualPlease read this Instruction Manual carefully before installing and operating the instrument
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.08410distance between the angle and the instrument is recommended.Do not mount pressure regulators dire
Bronkhorst High-Tech B.V.EL-FLOW Prestige 119.17.0842.7 Analog / digital operation2.7.1 Analog / local operationConnect the EL-FLOW Prestige to the po
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.084122.9Zeroing (using micro switch)The zero-point of each instrument is factory adjusted. If so requir
Bronkhorst High-Tech B.V.EL-FLOW Prestige 139.17.0843 Basic operation3.1 Mass flow measurement and controlAfter correct installation of the EL-FLOW Pr
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.08414The multi-gas/multi-range functionality is available for the full temperature and pressure range f
Bronkhorst High-Tech B.V.EL-FLOW Prestige 159.17.084Via the 9-pin sub-D side connector the instrument can be operated by means of:Analog interface (se
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.084163.3Analog operationThe following analog signals are available for each instrument through the 9-pi
Bronkhorst High-Tech B.V.EL-FLOW Prestige 179.17.0843.4.1 Hook-upA special T-part cable (7.03.366) is required for connecting the 9-pin sub-D side con
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.084183.4.2FlowDDERS232 communication can be used for operating the instrument using the Bronkhorst Flow
Bronkhorst High-Tech B.V.EL-FLOW Prestige 199.17.0843.4.3 Software (DDE applications)Examples of free Bronkhorst DDE client applications: FlowDDE, Flo
DisclaimerThe information in this manual has been reviewed and is believed to be wholly reliable. No responsibility, however, is assumed forinaccuraci
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.084203.5Basic RS485 (FLOW-BUS/Modbus) operationThis section is limited to RS485 FLOW-BUS or Modbus comm
Bronkhorst High-Tech B.V.EL-FLOW Prestige 219.17.0843.5.1Hook-upThe illustrations below show examples of a number of EL-FLOW Prestige instruments in a
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.084223.5.2SoftwareWhen using a pc to communicate with EL-FLOW Prestige instruments only the FLOW-BUS pr
Bronkhorst High-Tech B.V.EL-FLOW Prestige 239.17.0843.6 Other field bus configurationsThe following field buses are optionally available for the EL-FL
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.08424Changing PROFIBUS DP node addressThe node address can be easily set by using the rotary switches o
Bronkhorst High-Tech B.V.EL-FLOW Prestige 259.17.084 red 'Error' LEDTimeIndication OffContinuousNo error OnContinuousCritical error message
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.084263.8Micro switch functionsBy means of manual operation of the micro push-button switch some importa
Bronkhorst High-Tech B.V.EL-FLOW Prestige 279.17.0843.9Basic parameters and properties3.9.1 IntroductionMost instrument parameters can only be accesse
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.084283.9.2 Basic measurement and control parametersThe list below provides the most basic parameters fo
Bronkhorst High-Tech B.V.EL-FLOW Prestige 299.17.0843.9.4Basic alarm and counter settingsThe alarm and counter settings are most easily accessible via
WarrantyBronkhorst products are warranted against defects in material and workmanship for a period of three years from the date ofshipment provided th
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.08430FlowDDEparameterParameter namePurposeSection116Alarm Maximum LimitAlarm settingsSection 4.2.4117Al
Bronkhorst High-Tech B.V.EL-FLOW Prestige 319.17.084FlowDDEparameterParameter namePurposeSection349Mix Fraction IndexFluidset propertySection 4.2.3350
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.084324 Advanced operation4.1 Sealing material compatibilityEL-FLOW Prestige instruments are equipped fr
Bronkhorst High-Tech B.V.EL-FLOW Prestige 339.17.0844.2 Advanced parameters and properties4.2.1 Advanced measurement and control parametersMeasured Va
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.08434ResetTypeAccessRangeFlowDDEFLOW-BUSModbusUnsigned charR0…5114115/80x0E68/3689This parameter is use
Bronkhorst High-Tech B.V.EL-FLOW Prestige 359.17.084ValueModeInstrument actionSetpoint sourceSlave factor20Valve SteeringSetpoint is redirected to ‘55
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.08436CapacityTypeAccessRangeFlowDDEFLOW-BUSModbusFloatRW 1E-10...1E+10211/130x8168...0x8169/33129...331
Bronkhorst High-Tech B.V.EL-FLOW Prestige 379.17.084Pressure InletTypeAccessRangeFlowDDEFLOW-BUSModbusFloatRW 0...3.4E+38178113/130xF168...0xF169/6180
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.08438Mix Fraction PressureTypeAccessRangeFlowDDEFLOW-BUSModbusFloatRW 0...3.4E+38348126/60xFE30...0xFE3
Bronkhorst High-Tech B.V.EL-FLOW Prestige 399.17.0844.2.4Advanced alarm parametersBronkhorst digital instruments have a built-in alarm function. It is
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.0844Table of contents...
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.08440Alarm Delay TimeTypeAccessRangeFlowDDEFLOW-BUSModbusUnsigned charRW0…25518297/70x0C27/3112This val
Bronkhorst High-Tech B.V.EL-FLOW Prestige 419.17.0844.2.5Advanced counter parametersCounter ModeTypeAccessRangeFlowDDEFLOW-BUSModbusUnsigned charRW0..
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.084424.3Field bus operation4.3.1 FLOW-BUS master/slave controller operationEL-FLOW Prestige instruments
Bronkhorst High-Tech B.V.EL-FLOW Prestige 439.17.084Side connector RS485 communicationChange the baud rate, node address or parity of the side-connect
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.08444When the controller output is used for pump or external valve steering (Mass Flow Meters only), ma
Bronkhorst High-Tech B.V.EL-FLOW Prestige 459.17.084alternative pulse length must be specified on order.Provide a pull-up resistor of 5...10 kOhm to c
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.084464. Write the new value to parameter ‘86IOStatus’5. Set parameter ‘7Init Reset’ to 82Note: do not u
Bronkhorst High-Tech B.V.EL-FLOW Prestige 479.17.0845.2TroubleshootingIn this section a number of possible errors/malfunctions of the instrument are l
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.08448SymptomPossible causeActiontightness through valveNon-zero offset signal (occurs at upward/downwar
Bronkhorst High-Tech B.V.EL-FLOW Prestige 499.17.0846 Removal and return instructionsInstrument handlings:Purge gas linesWhen toxic or dangerous fluid
Bronkhorst High-Tech B.V.EL-FLOW Prestige 59.17.084...
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.08461 Scope of this manual1.1 IntroductionThis manual covers the EL-FLOW Prestige series mass flow mete
Bronkhorst High-Tech B.V.EL-FLOW Prestige 79.17.0841.4 References to other applicable documentsBasic instructionsDimensional drawingsDocument 9.17.083
Bronkhorst High-Tech B.V.EL-FLOW Prestige 9.17.08481.5 Model key
Bronkhorst High-Tech B.V.EL-FLOW Prestige 99.17.0842 Starting up2.1 Check functional propertiesBefore installing your EL-FLOW Prestige mass flow meter